ESTIMATION THE PHYSICAL ACTIVITY OF HEART FAILURE PATIENTS BASED ON TELEMETRY DATA

Author(s)

Melczer C1, Melczer L2, Raposa L. B1, Boncz I1, Molics B3, Ács P1
1University of Pécs, Pécs, Hungary, 2Heart Institute, Pécs, Hungary, 3University of Pécs, Pécs, BA, Hungary

OBJECTIVES: In our study we examined the physical activity of patient suffering from heart failure with Actigraph GT3X+ movement sensor simultaneously with the implanted CRT device to validate the internal sensors, and estimate the result of the physical performance reached in the 6-minute walking test (6MWT) based on the Patient Activity% (PA%).

METHODS: Biotronik CRT-P or CRT-D was implanted in all patients with heart failure included in the study. The 6MWT and the physical activity data from CRT-P or CRT-D and Actigraph were collected. The patients wore the Actigraph for a week (29.09.2018-10.10.2019), and the Home Monitoring data collection covered a year (01.07.2018-01.07.2019). SPSS 23 statistical package and Excel program were used for data processing. We used descriptive statistics, correlation analysis, Bland–Altman test and linear regression to evaluate the findings.

RESULTS: The Bland-Altman plots showed furthermore that there is correlation between the mean and the difference, with higher means resulting in larger (negative) differences (R2=0.109; F=4.665; p=0.037; Beta=-0.362; p=0.037) between the Actigraph GT3X+ and the built-in device activity datas. Compared the physical activity time spent in a day’s % from the Actigraph versus the results of 6MWT, where a moderate significant correlation among the variables (r=0.338; 0.031) was found. The analysis of the 6MWT and the PA% data revealed correlation with a linear regression (F= 6.126; p= 0.018). With this regression equation the assessing distance of the 6MWT can be estimated.

CONCLUSIONS: Based on the statistical analysis we found negative correlation in mean differences between the accelerometer PA% and built-in device’s PA%, which supports the fact that the physical activity was slightly overestimated by the built-in PA%. Based on this fact, further investigation is required about the CRT device's accelerometer and that of the PA%. The PA% result is applicable in estimating the walking distance of 6MWT.